Capter3
Function
Setting and Basic Operation
3-94
Curve Tracer CS-5000 Series Instruction Manual
Executing "RECALL DEFAULT"
[Procedure 2]
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F2 key on the right of the submenu "RECALL."
(3) Press F4 key on the right of the submenu "RECALL DEFAULT."
SETUP (setting conditions) is set to defaults.
For detailed DEFAULT settings, see "Table 3.11.1 Recall functions" in Section 3.11.8 "Default
settings."
[Procedure 3] Recalling TRACE data from USB memory to screen
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F2 key on the right of the submenu "RECALL."
(3) Press F1 key on the right of the submenu "SETUP/TRACE/REF" and select "TRACE."
(4) Press F3 key on the right of the submenu "FILE LIST."
SETUP information is displayed in the submenu "USB File list" and the waveform display area.
(5) Rotate FUNCTION knob or press F1(
↑
)/F2(
↓
) key to select the file.
(6) Press F6 key on the right of the submenu "SELECT."
The selected file name is displayed in the submenu "FILE LIST."
(7) Press F6 key on the right of the submenu "RECALL" and execute "RECALL."
(8) "Recall from ABCD0001.WFM" is displayed in the message area on the upper part of the
screen and TRACE waveform is displayed in the waveform display area.
At the same time, the menu screen disappears.
(3) (4)
(5) (6)
(7) (8)