COLLECTOR SUPPLY SETUP submenu
Pressing F3 function key allows COLLECTOR SUPPLY SETUP submenu on the right in the diagram below
to appear.
Key / knob
Setting items
Function description
F1
-
It displays the mode polarity (+/-) of COLLECTOR SUPPLY.
* For polarity setting, refer to section 3.7.2: MODE POLARITY key.
F2
FUNCTION
MINIMUM
Press F2 key on the right menu in the diagram above to select the
setting target.
It sets the minimum value of the sweep range of COLLECTOR
SUPPLY. If any value exceeding the setting value of MAXIMUM or
START VARIABLE below is set, the value of MAXIMUM or START
VARIABLE changes together; i.e. the same value is set.
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F3
FUNCTION
MAXIMUM
Press F3 key on the right menu in the diagram above to select the
setting target.
It sets the maximum value of the sweep range of COLLECTOR
SUPPLY. If any value less than setting value of MINIMUM or START
VARIABLE is set, the value of MINIMUM or START VARIABLE
changes together; i.e. the same value is set.
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F4
FUNCTION
RESOLUTION
Press F4 key on the right menu in the diagram above to select the
setting target.
It sets the sweep resolution of COLLECTOR SUPPLY. If operation is
done by this option, the resolution of this setting is used for sweep.
- Setting range: 0.1 to 5.0%
- Setting resolution: 0.1%
F5
FUNCTION
START VARIABLE
Press F5 key on the right menu in the diagram above to select the
setting target
It sets the sweep START value of COLLECTOR SUPPLY. Make setting
between MAXIMUM and MINIMUM values above.
If not, the setting value stops at MAXIMUM or MINIMUM value; i.e.
setting cannot be made out of the range.
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F6
-
(Not used)