Capter3
Function
Setting and Basic Operation
3-32
Curve Tracer CS-5000 Series Instruction Manual
3.7.3 POWER
−
WATT key
3.7.3.1
POWER-WATT key and display menu
It can limit the maximum peak power to be applied to a measured device; i.e. protecting the
device from breakage.
Pressing POWER - WATT key
③
allows "MAX PEAK POWER WATTS" menu to appear.
High Voltage mode HC(CS-5200
/
CS-5300)
HC(CS-5400)
3.7.3.2
Submenus and functions
This setting value and the maximum peak voltage setting value determine the value of the
current limitation resistance internally connected to the collector output in series. See Table 3.7
POWER WATTS/VOLTAGE and the current limitation resistance in Section 3.7.
In addition, if the maximum peak voltage setting is 3 kV, 390 W cannot be selected.
High Voltage mode
High Current mode
F1 Sets the maximum power to be applied to a
measured device
- When the maximum peak voltage is 30V,
and 300V, one of the power is selected by
390W, 120W, 12W, 1.2W., and 120mW.
- When the maximum peak voltage is 5kV, ,
one of the power is selected by 32W,
3.2W., and 320mW.
Sets the maximum power (current) to be applied to
a measured device. Following one is selected by
the model.
- CS-5200 : 4Kw, 400W
- CS-5300 : 10kW, 4kW, 400W
- CS-5400 : 12kW, 4.5kW, 450W
F2
F3
F4
(Not used)
F5
F6 (Not used)
* 1 The HIGH CURRENT mode is not installed in CS-5100.
* 2 10kW cannot be set for CS-5200.
F1
F2
F3
F4
F5
F6