Capter3
Function
Setting and Basic Operation
3-74
Curve Tracer CS-5000 Series Instruction Manual
Executing SYSTEM INITIALIZE
[Procedure]
(1) Press F2 key. The confirmation screen (Do you initialize system? Push [OK] button to initialize.
Push [CANCEL] button to abort.) appears.
(2) Press [OK] button. The confirmation screen (System initialization completed. Remote I/F setup
changed. Reboot for new remote setup.) appears.
(3) Restart this instrument using Power On/Standby switch.
Caution!
Execution of SYSTEM INITIALIZE initializes all internal settings. Necessary data of SETUP (setting
conditions), TRACE (waveform data), and REF (reference waveform data) should be saved in the USB
memory using SAVE function (see Sections 3.11.3 and 3.11.4) in SAVE/RECALL menu. For initialized
items, see the default setting items in "Table 3.11.1: Recall function" and "Table 3.11.2 Not recall
function" of Section 3.11.8.