Capter3
Function
Setting and Basic Operation
Curve Tracer CS-5000 Series Instruction Manual
3-15
Caution
!
At HIGH VOLTAGE MODE,
connect Collector terminal using
RED color wire which comes with
this unit as standard accessory
and carefully avoid incorrect wiring
These connections may cause of
damage on DUT and/or
malfunction of the unit.
3.3.9 Connecting the measured device with the test adapter CS-508 (Option)
[Connection procedure]
(1) Open the cover of the test fixture.
If the cover is opened, this instrument is changed to INTERLOCK state and its output is shut off.
(2) Connect the test adapter CS-508 with the measured device.
1
)
Pull the knob of
②
-B to the right, and set the measurement device in
②
-A.
2
)
Loosen a fixed screw of
②
-C, pull the knob of
②
-C to the left, and expand the applied part of
②
-A
when you cannot set the measurement device in
②
-A only by the knob of
②
-B.
Tighten a fixed screw of
②
-C and fix when you can attach the measurement device.
For diode: Connect the anode and cathode to the collector and emitter of the test adapter.
(3) Mount the test adapter on the test fixture.
(4) When mounting the test adapter, push it into the text fixture vertically while keeping its rear
surface parallel to the mounting part of the patch panel.
(5) When the cover of the test fixture is closed after connecting the test adapter, INTERLOCK state
is released and measurement can be carried out.
④
-1
④
-3
②
-A
②
-B
②
-C
④
-2
④
-4
③
Screws for the fixation
Test fixture
Example of shape of
device to be measured
* Attach in the following direction.
Left toward the applied part of
②
-A: ANODE
Right toward the applied part of
②
-A: CATHODE