Capter3
Function
Setting and Basic Operation
Curve Tracer CS-5000 Series Instruction Manual
3-31
3.7.2.3
Items related to step generator and vertical axis source
The collector supply MODE setting has the relationship with the step generator and the vertical
axis source as shown below:
The voltage of the waveform being set is generated to the collector or base.
HIGH VOLTAGE
HIGH VOLTAGE
Step generator
STEP/PULSE
Vertical axis source
AC
Changes to STEP
(PULSE is also selectable.)
COLLECTOR: collector current
SINE
(full-wave rectified wave)
Changes to STEP
COLLECTOR: collector current
DC
Changes to PULSE
COLLECTOR: collector current
LEAKAGE
Changes to DC
No. of steps fixed to 0
EMITTER: emitter current
(1000 times of collector current sense)
Example of switchover of the collector supply from RECTIFIED SINE (full-wave rectified
wave) to LEAKAGE
Items enclosed by
○
below are changed.
HIGH CURRENT
•
The HIGH CURRENT mode is not installed in CS-5100.
HIGH CURRENT
STEP/PULSE setting
Vertical axis source display
•
HC
Temporarily changes to PULSE.
COLLECTOR: collector current
Configuration for HIGH CURRENT
Collector/StepGen/Common
fixed
Leak icon
Grounding
point
DC icon