Capter4
Characteristics
Measurement
Curve Tracer CS-5000 Series Instruction Manual
4-35
In connection diagram, connected procedure, and the measurement procedure of this example, it
explains by the example that uses CS-5400 as the main unit, CS-304 as the test fixture.
Connection diagram
The diagram below shows how to directly connect a measured device to the test fixture.
Connect the gate, emitter, and collector terminals of IGBT to the base, emitter, and collector
terminals of the test fixture using the attached wires.
Connection procedure
(1) Remove the test adapter mounting part from the test fixture.
(2) Set the measured device IGBT in it.
(3) Connect the collector of the measured device IGBT to SENCE terminal, the emitter to SENCE
terminal, and the gate terminal to 300Vpk, using the attached wires as shown below:
Caution!
If SENSE of HIGH CURRENT is not connected, CS-5000 displays and measures nothing in HIGH
CURRENT mode.
To improve accuracy, connect the emitter sense.
Test wire for
HIGH CURRENT
Remove the patch panel.
Collector Emitter
Gate
Measured device IGBT