Chapter 6 Performance
6-4
Curve Tracer CS-5000 Series Instruction Manual
(4/4)
Items
Product specifications
Body
Outside dimensions (W
×
H
×
D) 424 mm
×
220 mm
×
555 mm (CS-5100)
424 mm
×
354 mm
×
555 mm (CS-5200/5300/5400)
(Excluding attachment and projection)
Mass
About 28 kg (CS-5100, excluding attachment and options)
About 43 kg (CS-5200/5300/5400, excluding attachment
and options)
Environment
conditions
Performance ensured
temperature range
+
10
°
C to
+
35
°
C
Operation ensured
temperature range
0
°
C to
+
40
°
C
Operation ensured
humidity range
0 to
+
40
°
C
5
%
to 80
%
RH (at 30
°
C or less) and no dew condensation
allowed
Upper limit value: 55
%
RH (at 40
°
C) and no dew
condensation allowed
Warming up time
30 minutes or more after powering on
Note) The specifications standards are guarantee values
after 30 minutes or more from turning on the power
supply pass.
Storage temperature range
Storage humidity range
−
20
°
C to
+
60
°
C, 5
%
to 80
%
RH (without dew
condensation)
Altitude (air pressure)
At operation: 2000 m or less (air pressure: about 79 kPa)
At non-operation: 15000 m or less (air pressure: about 12
kPa)
Pollution degree
2
Power supply environment
IEC60364-4-443: For protection Against Overvoltage’s of
Atmospheric Origin or Due to Switching
Power supply transient maximum voltage does not
exceed IEC60364-4-443 CAT II.
Temporary Overvoltages
Short-term: 1440 V
Long-term: 490 V
Accessories
Test fixture :1
- CS-303
:
CS-5100/5200/5300
- CS-304
:
CS-5400
Test adapter CS-500 (Blank adapter) :1
Wire set
:1 set (7 wires)
Instruction manual (Printed matter)
:1
CD (Instruction manual / Remote control manual) :1
Power cord
:1
Cord strap
:1