Wear part replacement
Take out valve sets from the pump head.
Clean the pump head as necessary.
Take apart valve sets and replace wear parts with new ones as necessary. Reassemble
and remount the valve sets in the pump head.
Observe the mounting order and direction of valve set components. See the exploded view at each
model.
NOTE
• Observe the mounting direction of the valve set.
• The valve seat must be held in the valve guide by
the seat holder when assembling an IX-060 TC/
TE valve set. Do not push the valve seat down
into the valve guide. Otherwise, a leak may result.
Connect pipes and the pump.
For the IX-C060 TC/TE type, tighten the adapters (16) to 4.4N•m. Excessive force may break the plastic
parts.
Diaphragm replacement
Necessary tools
• 3mm hexagon wrench • Spanner (13mm for C150/ 8mm for C060) • Torque wrench
Remove pipes from the pump.
NOTE
• The valve set may come down as the suction pipe is removed.
Take care not to drop it.
• Be careful not to get wet with residual chemicals in the pump
head or the discharge pipe.
1
3
4
2
Valve set
Valve
guide
Seat
holder
Valve
seat
Valve
O ring
Do not push
the valve seat.