Power Measurement Commands
Copyright © Itech Electronic Co., Ltd.
25
Chapter5 Power Measurement Commands
FETCh[:SCALar]:POWer:ACTive?
This command is used to read the latest pre-processing active power.It will not
affect the device setting values and will not trigger a new measurement
operation.It only requires to get the latest available value that has been
measured.
Command Syntax
FETCh[:SCALar]:POWer:ACTive?
Returned value
<NR2>
MEASure[:SCALar]:POWer:ACTive?
This command will trigger the measurement operation to get a real-time active
power.
Command Syntax
MEASure[:SCALar]:POWer:ACTive?
Returned value
<NR2>
FETCh[:SCALar]:POWer:APParent?
This command is used to read the latest pre-processing apparent power.It will
not affect the device setting values and will not trigger a new measurement
operation.It only requires to get the latest available value that has been
measured.
Command Syntax
FETCh[:SCALar]:POWer:APParent?
Returned value
<NR2>
MEASure[:SCALar]:POWer:APParent?
This command will trigger the measurement operation to get a real-time
apparent power.
Command Syntax
MEASure[:SCALar]:POWer:APParent?
Returned value
<NR2>