Calibration Commands
Copyright © Itech Electronic Co., Ltd.
73
CALibrate:CURRent[:LEVel]
This command only used for calibration mode.
It enters a calibration voltage value that you obtain by reading an external
meter.You must first select a calibration level (with CALibrate:VOLTage:POINt)
for the value being entered. These constants are not stored in nonvolatile
memory until they are saved with CALibrate:SAVE.
Command Syntax
CALibrate:CURRent[:LEVel] <NRf>
Parameters
external reading
Unit:A (amps)
Example:CAL:CURR 3.2223
Relevant command
CAL:STAT CAL:SAV
CALibrate:CURRent:EXTernal:POINt<point>
This command is used to calibrate external current points.
Command Syntax
CALibrate:CURRent:EXTernal:POINt<point>
Parameters
<point>
CALibrate:CURRent:EXTernal:[LEVel] <NRf>
This command is used to input external calibration current.
Command Syntax
CALibrate:CURRent:EXTernal:[LEVel] <NRf>
Parameters
<NRf>
CALibrate:VOLTage:POINt
This command can only be used in calibration mode. It is used to set the
calibration points of constanrd voltage mode. P1, P2 is used in low voltage
meter range, P3, P4 is used in high voltage meter range. It can use calibrate
voltage source and voltage meter.
Command Syntax
CALibrate:VOLTage:POINt <point>
Parameters
P1 | P2 | P3 | P4
Example:CAL:VOLT:POIN P2