28 of 36
REGLO ICC - ENGLISH , REV. D, 14-036-E
18.2.4 Set direction
Set direction to clockwise:
*
1J[CR]
REQUEST
RESPONSE
18.3 Modes and Settings
18.3.1 Operational pumping mode
Get the current mode of channel 1.
L[CR][LF]
1xM[CR]
REQUEST
RESPONSE
18.3.2 Pumping rate (drive speed)
Set drive speed of channel 2 to 123 RPM:
*
2S012300[CR]
REQUEST
RESPONSE
18.3.3 Pumping rate (volume/time)
Set pumping rate of channel 1 to 1.3 μL/min:
1300E-3[CR][LF]
1f1300-3[CR]
REQUEST
RESPONSE
18.3.4 Volume
Set the volume of channel 4 to 300 mL:
3000E+2[CR][LF]
2[CR]
REQUEST
RESPONSE
The current mode of channel 1 is RPM.
Set the mode of channel 4 to Time.
*
4N[CR]
REQUEST
RESPONSE
Get the current drive speed of channel 4:
12.30[CR][LF]
4S[CR]
REQUEST
RESPONSE
The current drive speed of channel 4 is 12.3 RPM.
Get current pumping rate of channel 4:
3250E+1[CR][LF]
4f[CR]
REQUEST
RESPONSE
The current pumping rate of channel 4 is 32.5 mL/min.
Get the volume of channel 2:
1300E-3[CR][LF]
2v[CR]
REQUEST
RESPONSE
The volume of channel 3 is 1.3 μL.