6.4.3 SETUP Menu
In the SETUP Menu the tubing ID is set by selecting the size of tubing being used in the
highlighted channel selected in the upper right hand corner. The tubing ID is selected from
the drop down menu. The number of roller backsteps for drip free dispensing can also be set
between 0–100 (0 = Default Value).
6.4.4 CALIBRATION Menu
In the CALIBRATION Menu each channel can be calibrated independently. The channel being
calibrated is listed in the upper right hand corner.
The Reglo ICC allows each channel to be set up and calibrated. This allows for maximum
fl exibility and optimum channel to channel precision. Prior to the calibration each channel must
be confi gured for tubing ID, the number of backsteps that are desired for drip free dispensing,
and calibration of that channel.
The Reglo ICC allows you to enter the volume of fl uid to dispense for calibration, dispense
time for the specifi ed volume, and pumping direction. Once you have entered those values,
dispense desired volume into a container. Measure the actual amount of the fl uid by weight
or volume and enter that value on the last screen to adjust the pump for your application. The
calibration can be interrupted by pressing the Run/Stop button which will cancel the existing
progress. New uninterrupted calibration run will be required in order to calibrate the channel
properly. When adjusting the dispense time for calibration, if requested time is shorter than
the pump’s physical capabilities, the dispense time will automatically default to minimum
achievable dispense time.