You can complete the calibration by pressing OK/Enter button following your measurement
input. You will receive a message and checkmark once your calibration is successful.
You can cancel the calibration after the calibration run by pressing the Reset button. You will
receive the message displaying that the calibration has been canceled.
6.4.5 GLOBAL SETTINGS Menu
In the GLOBAL SETTINGS Menu the display language can be selected from English, French,
Spanish, and German. The pump can also be reset to the factory defaults at this menu.
6.5 Installing Tubing
1.
Switch the pump OFF.
2.
Remove the cassette by slightly pressing the fixing-tongue and lifting it simultaneously.
3.
Insert the 2-stop or 3-stop color-coded tubing with one stopper into the cassette.
4.
Let the tubing hang down (prevents it from twisting).
5.
Insert the tubing with the second stopper at the other end of the cassette.
6.
Reinsert the cassette into the roller-head.
7.
Make sure each cassette is positioned and aligned well between
dedicated segments on the bars.
8.
Turn the pump ON.
9.
When the pump is idle, release all cassettes to maximize tubing life
(beware of siphoning or back flow).