Reglo ICC

Operating Manual

Summary of Contents for EW-78001-70

Page 1: ...Reglo ICC Operating Manual ...

Page 2: ...2 of 36 REGLO ICC ENGLISH REV D 14 036 E Reglo ICC Operating Manual Download Pump Control Software www masterflex com ...

Page 3: ...sponse Messages 20 15 1 Status Response 20 15 2 Data Response 20 16 Request Messages 21 16 1 Set Pump Address 21 16 2 Device Command 21 16 3 Tube Diameter Index 25 Table of Contents 1 Safety Precautions 5 2 Warranty 5 3 Product 6 4 Rear Panel 6 5 Main Voltage 6 6 Pump Operation 7 6 1 Key Pad and Screen 7 6 2 Icons 7 6 3 Changing Parameters 8 6 4 Menus 8 6 4 1 STATUS Menu 8 6 4 2 PUMPING Menu 9 6 4...

Page 4: ...18 1 2 Event message enable 27 18 2 Pump Drive 27 18 2 1 Start pumping 27 18 2 2 Stop pumping 27 18 2 3 Get direction 27 18 2 4 Set direction 28 18 3 Modes and Settings 28 18 3 1 Operational pumping mode 28 18 3 2 Pumping rate drive speed 28 18 3 3 Pumping rate volume time 28 18 3 4 Volume 28 18 3 5 Run time 29 18 3 6 Pause time 29 18 3 7 Cycle Count 29 18 4 Configuration 29 18 4 1 Tubing Inside Di...

Page 5: ...sures based on the specific situation must be taken When pumping corrosive media we recommend placing the pump in a collecting basin The pump should not come in contact with liquid Do not open or remove the housing while the pump is operating Service and repairs carried out by the customer or by third party companies voids all warranties Appliance coupler of the power supply is used to disconnect t...

Page 6: ...al ISMATEC representative immediately 4 Rear Panel 1 24 V DC power Inlet plug 2 On Off Switch circuit breaker 3 RS 232 In DB9 female 4 RS 232 Out DB9 male 5 USB 2 0 Mini B 5 1 3 4 2 5 Main Voltage Use only the supplied power supply and cord for your Reglo ICC pump DO NOT use any other substitutes or damage may occur The power supply is designed to provide DC voltage for your Reglo ICC pump It has ...

Page 7: ...thout changing b Resets dispensing cycle when pump is paused 3 Directional Arrows a Used to navigate screen and highlight selections b Up down left right 4 OK Enter Button a Press to edit parameter b Saves edit to parameter once complete 5 LCD Display 5 4 2 3 1 6 2 Icons 1 2 3 4 5 6 7 1 STATUS Menu 2 PUMPING Menu 3 SETUP Menu 4 CALIBRATION Menu 5 GLOBAL SETTINGS Menu 6 Channel Number Next Screen 7...

Page 8: ... stop all the channels When the pump is idle pressing the Run Stop button will start pumping to the settings for each channel When the pump is running pressing the Run Stop button will stop all channels If a timed or repeating pumping mode is being used pressing the Run Stop button will pause the program Pressing the Run Stop button again will resume the program The Reset button is used to cancel ...

Page 9: ... are rotation direction volume volume units time and time units Rotation direction can be selected as CW or CCW Combinations of time and volume can be set to any value as long as it is in the range of the pump s capabilities If the time value entered is less than what the pump is capable of for desired volume it will default to the shortest dispense time possible Volume units selectable are μL mL ...

Page 10: ...g ID see Section 6 4 3 Selectable flow rate units are RPM μL min mL min and L min Selectable volume units are μL mL and L Selectable pause time units are sec min and hours Only flow rate can be changed during pumping or when the operation is paused Dispensing program can be reset during operation by pressing the Reset button after pumping is paused 6 4 2 6 Time with Pause Mode Time with Pause Mode a...

Page 11: ... channel must be configured for tubing ID the number of backsteps that are desired for drip free dispensing and calibration of that channel The Reglo ICC allows you to enter the volume of fluid to dispense for calibration dispense time for the specified volume and pumping direction Once you have entered those values dispense desired volume into a container Measure the actual amount of the fluid by wei...

Page 12: ...panish and German The pump can also be reset to the factory defaults at this menu 6 5 Installing Tubing 1 Switch the pump OFF 2 Remove the cassette by slightly pressing the fixing tongue and lifting it simultaneously 3 Insert the 2 stop or 3 stop color coded tubing with one stopper into the cassette 4 Let the tubing hang down prevents it from twisting 5 Insert the tubing with the second stopper at...

Page 13: ...elps to protect the tubing from unnecessary strain and prolongs its service life However a siphoning effect can occur when the tubing cassette is released from the pump head Use caution to prevent the fluid from flowing back to the reservoir 7 Over Current Protection The Reglo ICC features an external power supply that has overload protection as well as a combined circuit breaker and On Off switch o...

Page 14: ...e Due to the configuration of the Reglo ICC Pump the use of pressure lever cassettes is not recommended 9 Repair The Reglo ICC is designed for many years of trouble free service The Reglo ICC does not have any fuses or adjustments inside the pump If you wish to attempt any repairs after the warranty period please contact your local ISMATEC Distributor for parts 10 Cleaning When replacing the tubes...

Page 15: ...justable in steps of 0 01 rpm Differential pressure Max 1 0 bar 14 5 psi Remote control Via USB 2 0 or RS 232 digital interfaces Main power connection 100 240VAC 50 60 Hz Power consumption Max 30 W Operating conditions Temperature 5 to 40 C 41 to 104 F Rel humidity max 80 Non condensing at normal laboratory conditions Dimensions Weight D x W x H 205 x 125 x 170 mm 8 1 x 5 x 6 7 Weight 2 7 kg 6 lbs...

Page 16: ...063 6 3 1 09 SC0060T 0 072 7 2 1 14 SC0061T 0 078 7 8 1 22 SC0062T 0 088 8 8 1 3 SC0063T 0 10 10 1 42 SC0064T 0 11 11 1 52 SC0065T 0 13 13 1 65 SC0066T 0 15 15 1 75 SC0067T 0 16 16 1 85 SC0068T 0 17 17 2 06 SC0069T 0 20 20 2 29 SC0070T 0 24 24 2 54 SC0071T 0 27 27 2 79 SC0072T 0 31 31 3 17 SC0224T 0 35 35 Cassettes for the ICC pumps are designed to work with stoppered tubing that fits the MS CA Cas...

Page 17: ...14 4 Addressing The Reglo Digital protocol uses an addressing scheme where each pump is assigned a singular address This addressing is designed to accommodate multiple pumps which have been daisy chained to the same RS 232 bus The addressing scheme for the protocol described in the document is dependent on the physical layer used for communication This arrangement allows for the same command defini...

Page 18: ...rn 0x0d LF line feed 0x0a SP space 0x20 VB vertical bar or pipe 0x7c 14 6 Data Type Formats The following data formats are used in the subsequent message definitions 14 6 1 Boolean Width 1 Format A single character indicating True or False 0 False 1 True 14 6 2 Direction Width 1 Format A single character indication direction J Clockwise K Counter clockwise 14 6 3 Discrete Type 1 Width 1 to 4 Unit 1...

Page 19: ... x 10se For example 1 200 x 10 2 is represented with 1200E 2 Note that the decimal point is inferred after the first character mmmm 4 character mantissa E E character s Sign for exponent or e Single digit exponent 14 6 11 Time Type 1 Width 1 to 8 Units 0 1 second Range 0 to 35 964 000 0 to 999 hr Format Up to eight characters representing the time in 0 to 999 hours with a resolution of 0 1 sec Unu...

Page 20: ...some commands with different meanings for each See individual direct addressed command definitions for context 0x2b A positive response used for some commands with different meanings for each See individual direct addressed command definitions for context 15 2 Data response A data response is sent when parameter values must be provided in response to a request message Each data response is compose...

Page 21: ...ssage Device Command Fields Char Width 1 Variable Variable optional 1 optional Variable optional 1 Variable optional 1 1 VB VB Data Parameter 1 Command Code CR LF Data Parameter n Data Parameter 2 Addr Addr The address of a pump channel 1 to 4 For commands which interact with parameters on a per pump and not per channel basis this field is ignored The field must still be provided however In this c...

Page 22: ...ter 2 Get Parameter 1 C Cycle count of 0 R Max flow rate exceeded or flow is set to 0 V Max volume exceeded Parameter 2 Limiting value C Value is undefined R Max flow mL min V Max vol mL 3 0 OPERATIONAL MODES AND SETTINGS 3 1 xM Get the current channel or pump mode L RPM M Flow Rate O Volume at Rate G Volume over Time Q Volume Pause N Time P Time Pause 3 2 L Set pump channel to RPM mode Set 3 3 M ...

Page 23: ... configurable data to default values Set 5 0 CALIBRATION 5 1 xR Get direction flow for calibration Get DIRECTION J CW K CCW 5 2 xR Set direction flow for calibration J or K using DIRECTION format Set 5 3 xU Get the target volume to pump for calibrating mL Get Volume Type 1 5 4 xU Set the target volume to pump for calibrating using Volume Type 2 Set Volume Type 1 5 5 xV Set actual volume measured d...

Page 24: ...me value To individually set roller counts for each channel use the non legacy command designed for this operation Discrete Type 2 Set 6 25 V Get the current setting for pump time in 1 10 second Get Discrete Type 1 6 26 V Set the current setting for pump time in 1 10 second Discrete Type 2 Set 6 27 VM Set the current run time setting for dispensing in minutes Discrete Type 5 Set 6 28 VH Set the cu...

Page 25: ...Variable optional 1 1 VB VB Data Parameter 1 Event Code CR LF Data Parameter n Data Parameter 2 Event Indicator Event indicator A single caret character ASCII 0x53 which identifies this message as an event message Event code A string which identifies a specific type of event Data Parameter N A data value associated with the event There may be none or several of these These will vary by event type ...

Page 26: ... that the pump is alive and responsive It is also responsible for providing updates on the state of an in progress pumping operation Event Code U Data Parameter 1 Channel Description The channel associated with this status update event message Range 1 8 Format A single character representing the channel number Data Parameter 2 Status Description The current status of this channel Format A single c...

Page 27: ...le Get the current enable state of event messages 0 CR LF 0xE CR REQUEST RESPONSE Event messages are disabled 18 2 Pump Drive 18 2 1 Start pumping Start pumping on channel 2 2H CR REQUEST RESPONSE 18 2 2 Stop pumping Stop pumping on channel 3 3I CR REQUEST RESPONSE 18 2 3 Get direction Get direction of channel 1 K CR LF 1xD CR REQUEST RESPONSE The direction of channel 1 is counter clockwise 2 CR L...

Page 28: ... to 1 3 μL min 1300E 3 CR LF 1f1300 3 CR REQUEST RESPONSE 18 3 4 Volume Set the volume of channel 4 to 300 mL 3000E 2 CR LF 4v3000 2 CR REQUEST RESPONSE The current mode of channel 1 is RPM Set the mode of channel 4 to Time 4N CR REQUEST RESPONSE Get the current drive speed of channel 4 12 30 CR LF 4S CR REQUEST RESPONSE The current drive speed of channel 4 is 12 3 RPM Get current pumping rate of ...

Page 29: ...ST RESPONSE 18 4 2 Backsteps Set the channel 1 backsteps to 0 1 0000 CR REQUEST RESPONSE Get the run time of channel 2 20 CR LF 2xT CR REQUEST RESPONSE The current run time of channel 2 is 20 seconds Get the pause time of channel 2 20 CR LF 2xP CR REQUEST RESPONSE The current pause time of channel 2 is 20 seconds Get the cycle count of channel 2 20 CR LF 2 CR REQUEST RESPONSE The cycle count of ch...

Page 30: ...on Get the time since channel 3 was last calibrated 03596400 CR LF 3xX CR REQUEST RESPONSE Channel 3 was last calibrated 3 596 400 seconds 999 hours ago 18 5 5 Start calibration Start calibration on channel 2 2xY CR REQUEST RESPONSE Get the target calibration volume for channel 3 1000E 0 CR LF 3xU CR REQUEST RESPONSE The target calibration volume for channel 3 is 1 mL Get the measured calibration ...

Page 31: ...t pump s temporary display name Set the pump s display name to Reagent A 0xNReagent A CR REQUEST RESPONSE 18 6 3 Serial number Set the pump s serial number to 12345ABC 0xS CR REQUEST RESPONSE 18 6 4 Language Set the pump language to German 0xL3 CR REQUEST RESPONSE Get the pump s serial number 12345ABC CR LF 0xS12345ABC CR REQUEST RESPONSE Get the current pump language setting 0 CR LF 0xL CR REQUES...

Page 32: ...lution count for channel 3 is 1511 18 6 8 Volume total Get the total volume count for channel 3 0000001511 CR LF 3xG CR REQUEST RESPONSE The total volume count for channel 3 is 1511 mL 18 6 9 Time total Get the total pumping time count for channel 3 0000001511 CR LF 3xJ CR REQUEST RESPONSE The total pumping time count for channel 3 is 1511 seconds Get the pump channel count 3 CR LF 0xA CR REQUEST ...

Page 33: ...LF EVENT A pump complete event occurred on channel 2 18 7 2 Channel status update U3 A 0000000000 0000100000 0000 CR LF EVENT Channel 3 is currently pumping This channel is probably in volume mode as volume is the only provided field here There are 100 mL left to pump in the operation ...

Page 34: ... 61010 1 UL 61010 1 CAN CSA C22 2 No 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use Conforms to UL Std 61010 1 Certified to CAN CSA C22 2 Std No 61010 1 CAN CSA C22 2 No 61010 1 Tested to the requirements of CAN CSA C22 2 No 61010 1 second edition including amendment 1or a later version of the same testing requirements 2002 96 EC EC Directive Waste ...

Page 35: ...ected to other products recommended or described in the user documentation and Used in the same state as it was delivered from ISMATEC except for alterations described in the user documentation The Declaration of Conformity is valid only for systems that are marked with the CE marking Equipment connected to the Reglo ICC should meet the safety requirements of EN 61010 1 IEC 61010 1 or relevant har...

Page 36: ...act US Canada only Toll Free 1 800 MASTERFLEX 1 847 637 3739 Outside US Canada 1 847 549 7600 1 847 381 7050 www masterflex com techinfo masterflex com EN809 Manufactured by Cole Parmer Instrument Company 28W092 Commercial Avenue Barrington IL 60010 ...