2 OVERVIEW
Part No 1176122
9
Invacare® Solo2™ Portable Oxygen Concentrator
Recycle
DO NOT use oil or grease
Keep dry in transport or storage
IPX1
Protected against dripping water in upright position (AC and DC power input)
IPX2
Protected against dripping water in upright and tilted positions (battery power only)
NO AP/APG
Not suitable for use in the presence of a flammable anaesthetic mixture
POWER INPUT:
24 VDC @ 7.5 Amp or 11-16 VDC @ 10.0 Amp max (12.6 VDC Nom.)
EXTERNAL POWER SUPPLY INPUT:
AC POWER SUPPLY:
DC POWER SUPPLY:
120 VAC, 50/60 Hz @ 2.5 amps, 230 VAC, 50/60 Hz @ 1.25 amps
11-16 VDC, (12.6 VDC Nom. @ 10.0 Amp max)
SOUND LEVEL:
< 40 dBA weighted @ 2 LPM continuous and all pulse settings (1-5)
ALTITUDE:
Up to 10,000 ft (3046 m) above sea level
Titration recommended for use above 10,000 ft (3046 m)
OXYGEN CONCENTRATION:*
*BASED ON AN ATMOSPHERIC PRESSURE OF
14.7 PSI
(101 KPA) AT 70°F (21°C) NOMINAL
87% to 95.6%, after initial warm-up period (approximately 5 minutes) at all flow rates
CONSERVER TRIGGER SENSITIVITY:
FACTORY SET - NO ADJUSTMENT
PRESSURE ACTIVATED
<
0.20 cmH
2
O max pressure drop
All settings - using 7 ft (2.1 m) cannula