![Invacare Solo 2 TPO100B Internal Repair Manual Download Page 75](http://html1.mh-extra.com/html/invacare/solo-2-tpo100b/solo-2-tpo100b_internal-repair-manual_2082621075.webp)
26 LCD SCREEN NAVIGATION GUIDE
Part No 1176122
75
Invacare® Solo2™ Portable Oxygen Concentrator
26.6 Mode and Flow Adjust Screen
The Mode and Flow Adjust Screen allows the user to change the current mode of operation and output flow.
26.7 Changing Mode
1.
With the Mode & Flow screen displayed, press and hold the Mode/Return/Highlight Button for one second.
2.
Wait until the Mode & Flow screen flashes three times.
26.8 Adjusting Flow
1.
With the Mode & Flow screen displayed, press and hold the Up/Increase Button or Down/Decrease Button until desired setting is achieved.
2.
Wait until the Mode & Flow screen flashes three times.
To adjust either the mode of flow, the Mode and Flow Adjust Screen must be displayed. If the battery level screen is displayed, press
the Mode/Return/Highlight, Up/Increase or Down/Decrease button for approximately two seconds (screen should change to Mode
and Flow Adjust).
Any changes to the setting will not take effect until after the screen flashes three times.
Up/Increase Butto
Mode/Return/
Highlight Button
Down/Decrease But