6 ADMINISTRATIVE SETTINGS
Part No 1176122
21
Invacare® Solo2™ Portable Oxygen Concentrator
5.
Press and hold the Return/Highlight button to return to the main menu operational screen.
6.4
Toggling Auto Pulse
1.
Enter administrative settings screen (DETAIL "B"). Refer to Accessing Administrative Settings Screen on page 19 for this procedure.
2.
Locate the Auto Pulse setting (DETAIL "B").
A. Press the Up/Increase or Down/Decrease button to highlight Auto Pulse setting.
B.
Press the Return/Highlight button to relocate highlight bar to the adjustable setting on right side of screen.
3.
Press the Up/Increase or Down/Decrease button to toggle between ON or OFF.
4.
Save changes by pressing the Return/Highlight button to relocate highlight bar to left side of screen.
5.
Press and hold the Return/Highlight button to return to the main menu operational screen.
The portable oxygen concentrator will automatically return to the main menu operational screen if idel for a period of approximately 10-15
seconds.
The portable oxygen concentrator is preset by Invacare at a maximum setting of 2.0 LPM.
Changes are not saved until the highlight bar is relocated to the left side of the screen.
The portable oxygen concentrator will automatically return to the main menu operational screen if idel for a period of approximately 10-15
seconds.
Unit will remain in Auto Pulse mode until Auto Pulse is returned to “OFF”.