6 ADMINISTRATIVE SETTINGS
Invacare® Solo2™ Portable Oxygen Concentrator
20
Part No 1176122
6.2
Resetting Patient Hours
1.
Access the administrative settings screen. Refer to Accessing Administrative Settings Screen on page 19 for this procedure.
2.
Locate the Patient Hours setting (Detail "B").
A. Press Up/Increase or Down/Decrease button to highlight Patient Hours setting.
B.
Press the Return/Highlight button to relocate highlight bar to the adjustable setting on right side of screen.
3.
Press Up/Increase or Down/Decrease button to reset hours.
4.
Save changes by pressing the Return/Highlight button to relocate highlight bar to left side of screen.
5.
Press and hold the Return/Highlight button to return to the main menu operational screen.
6.3
Adjusting DC Power Limit
1.
Enter administrative settings screen (DETAIL "B"). Refer to Accessing Administrative Settings Screen on page 19 for this procedure.
2.
Locate the DC Power Limit setting (DETAIL "B").
A. Press the Up/Increase or Down/Decrease button to highlight DC Power Limit setting.
B.
Press the Return/Highlight button to relocate highlight bar to the adjustable setting on right side of screen.
3.
Press the Up/Increase or Down/Decrease button to adjust Power Limit.
4.
Save changes by pressing the Return/Highlight button to relocate highlight bar to the left side of the screen.
Changes are not saved until the highlight bar is relocated to the left side of the screen.
The portable oxygen concentrator will automatically return to the main menu operational screen if idel for a period of approximately 10-15
seconds.
Maximum DC Power Limit is 2.0 LPM continuous.
Changes are not saved until the highlight bar is relocated to the left side of the screen.