21 INLET RESONATOR
Invacare® Solo2™ Portable Oxygen Concentrator
64
Part No 1176122
21 Inlet Resonator
21.1 Replacing the Battery Input Board
1.
Disassemble the Solo2. Refer to Accessing Internal Components on page 40.
2.
Lift the resonator out of the Solo2 unit.
3.
Disconnect the tube at the base of the resonator that feeds the compressor.
4.
Attach the new resonator to the tube coming from the compressor.
5.
Place the resonator back into the Solo2 feeding the tubing as show.
6.
Reassemble the unit and test to ensure Solo2 is operating correctly.
Tube attaches from
compressor here
Rear Shroud
Resonator
DETAIL “A”
DETAIL “B”
DETAIL “C”
To Inlet Filter
To Inlet Filter
From Compressor to Resonator