![Invacare Solo 2 TPO100B Internal Repair Manual Download Page 44](http://html1.mh-extra.com/html/invacare/solo-2-tpo100b/solo-2-tpo100b_internal-repair-manual_2082621044.webp)
13 BATTERY PACK
Invacare® Solo2™ Portable Oxygen Concentrator
44
Part No 1176122
13.3 Replacing Batteries in Battery Pack
Removing
•
Required tools: Torx® T10 screwdriver
1.
Remove the label from the back of the battery pack (Detail "A") to expose the screws (Detail "B").
2.
Remove and retain the screws. Remove the back cover.
3.
Unplug each battery from the bottom connector (Detail "C").
CAUTION
When disassembling the battery pack make sure that no exposed connectors touch each other, otherwise, shorts of the individual batterys
may occur.
If the battery pack does not charge or the Solo2 does not power up in the required time as stated in Product Parameters, replace
all battery packs.
If the battery label rips during removal, remove the serial number label and replace the battery label with a new one. Apply the serial
number label to the replacement label.
!
Battery Pack
Label
Serial Number
Label
Battery Pack with
Label removed
Screws
Batteries
Battery Pack
Connector
DETAIL “A”
DETAIL “B”
DETAIL “C”