9 REPLACING THE INLET FILTER
Part No 1176122
37
Invacare® Solo2™ Portable Oxygen Concentrator
9 Replacing the Inlet Filter
1.
Position the Solo2 unit flat on its back to locate the Inlet Filter access filter door.
2.
Using a #2 Phillips® screwdriver remove the retaining screw securing the inlet filter access door to unit.
3.
Remove the filter access door from the unit.
4.
Locate the existing inlet filter and remove.
5.
Install the replacement inlet filter.
6.
Reinstall inlet filter access door by inserting tabs into slots located on the unit.
7.
Reinstall retaining screw to secure the inlet filter access filter door.
The inlet filter should be replaced per the Preventative Maintenance Record.
Use Invacare P/N 1157083, Inlet Filter.
Filter Access Door
Screw
Inlet Filter