![Invacare Solo 2 TPO100B Internal Repair Manual Download Page 38](http://html1.mh-extra.com/html/invacare/solo-2-tpo100b/solo-2-tpo100b_internal-repair-manual_2082621038.webp)
10 REPLACING THE PATIENT OUTLET FILTER AND FILTER CONNECTOR
Invacare® Solo2™ Portable Oxygen Concentrator
38
Part No 1176122
10 Replacing the Patient Outlet Filter and Filter Connector
10.1 Replacing Patient Outlet Filter
1.
With the SOLO2 in the upright position, locate the patient outlet filter.
2.
Place fingers on patient outlet filter and turn counter clockwise to loosen.
3.
Remove patient outlet filter and o-ring from unit.
4.
Install new o-ring onto unit.
5.
Install new patient outlet filter onto unit.
6.
Hand tighten filter until secure.
The patient outlet filter should be replaced according to the Solo2 Preventative Maintenance Record.
Use Invacare P/N 1164372, Patient Outlet Filter Kit.
Inlet Filter
O-Ring
Patient Outlet Filter Removed