17 CONTROL PANEL
Invacare® Solo2™ Portable Oxygen Concentrator
60
Part No 1176122
17.2 Replacing Control Panel Label
1.
Peel the control panel label off of the top of the SOLO
2
concentrator.
2.
Clean the top of the unit with isopropyl alcohol and non-abrasive cloth or sponge. Remove all residual glue from the top of the unit.
3.
Allow the top surface of the unit to dry completely to ensure a good seal of the replacement label.
4.
Install the new control panel label.
5.
Press firmly on the perimeter surfaces of the label.
Invacare recommends using isopropyl alcohol to remove residual glue from the top of the unit. The use of any other type of household or
industrial cleaner will cause damage to the unit and should not be used.
CAUTION
DO NOT press directly on the LCD display when applying replacement control panel label otherwise damage to the LCD display may
occur.
!
Control Panel Label
Domed Buttons
Domed Buttons