![Invacare Solo 2 TPO100B Internal Repair Manual Download Page 62](http://html1.mh-extra.com/html/invacare/solo-2-tpo100b/solo-2-tpo100b_internal-repair-manual_2082621062.webp)
19 EXHAUST MUFFLER
Invacare® Solo2™ Portable Oxygen Concentrator
62
Part No 1176122
19 Exhaust Muffler
19.1 Replacing the Exhaust Muffler
1.
Disassemble the Solo2. Refer to Accessing Internal Components on page 40.
2.
Remove the P.C. Board.
3.
Remove the Sieve Bed Unit.
4.
Follow the instructions in Removing the Compressor. DO NOT perform STEPS 6, 9, 11, 12, 13, 14 or 15.
5.
Pull the muffler assembly out of the compressor box and disconnect from the tube.
6.
Install a replacement muffler and place into the compressor box.
CAUTION
Before handling the P.C. Board, ensure you are properly grounded to prevent static discharge and damage to the components of the P.C.
Board. A static strap MUST be worn and properly grounded using an alligator clip. Electrical conduit or a water pipe can bue used if a
ground is unavailable. Care should be taken to ensure that the alligator clip contacts bare metal surface.
!
Muffler
Tube from exhaust side of the Compressor