EN – ENGLISH
9
8
Phase
Time
(Min.)
Water
Temperature
Detergent and
Concentration
Pre-wash 1 4:00
Cold water
N/A
Enzyme
Wash
4:00
Hot water
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio 1:128)
Wash 1
10:00
60.0ºC
(140º F)
Neutral pH
detergent
(e.g. Renu-
Klenz™ , Steris,
Ratio 1:256)
Rinse 1
0:30
Hot water
N/A
Thermal
Rinse**
2:00
82.2ºC
(180º F)
N/A
** Optional phase for disinfection of components –
minimum water temperature as indicated or per worker
manufacturer specifications for the thermal rinse cycle.
NOTE
– Any deviation from this guideline could result in
damage to the equipment as well as improper cleaning
results.
Rinse with purified water. Do not perform if parameters
cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
Lubrication
It is important that metal movable parts be properly
lubricated to keep these parts functional.
Inspection:
Carefully inspect all parts after each cleaning to assure all
components are in proper working con-dition.
CAUTION:
Components should be returned for repair if wear
is noted.
The life expectancy of the MAYFIELD products is expected
to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
Recommended Action
Recommended Frequency
Return the device to
the Integra LifeSciences
Repairs department for
detailed inspection and
servicing.
Once / year
Request that Integra
NeuroSpecialists perform
routine inspections of the
device
Twice / year
In the absence of proper care and servicing of the device,
negative effects may be seen after repeated processing
over time which may lead to reduced performance.
Contact information: See the Service and Repair section
for contact information on how to return your device for
periodic servicing and to request periodic inspections.
See Inspection and/or Service notes section for routine
checks to be performed on the device.
NOTE
– Any serious incident that has occurred in relation
to the device for the user and/or the patient should
be reported to the manufacturer and the competent
authority of the member state in which the user and/or
patient is established.
Device Disposal
NOTE: Follow hospital procedures for disposal of this
device.
Summary of Contents for A1112
Page 2: ...2 This page is intentionally left blank ...
Page 3: ...3 MAYFIELD Infinity Support System A1112 Instruction Manual EN ENGLISH ...
Page 12: ...12 FR FRANÇAIS FR FRANÇAIS Système de support Infinity MAYFIELD A1112 Mode d emploi ...
Page 21: ...21 21 IT ITALIANO Sistema di supporto Infinity MAYFIELD A1112 Manuale di istruzioni ...
Page 30: ...DE DEUTSCH 30 DE DEUTSCH MAYFIELD Infinity Stützsystem A1112 Gebrauchsanleitung ...
Page 39: ...39 39 ES ESPAÑOL Sistema de soporte Infinity MAYFIELD A1112 Manual de instrucciones ...
Page 57: ...57 57 This page is intentionally left blank ...
Page 58: ...58 This page is intentionally left blank ...
Page 59: ...59 This page is intentionally left blank ...
Page 60: ...60 This page is intentionally left blank ...
Page 61: ...61 This page is intentionally left blank ...
Page 62: ...62 This page is intentionally left blank ...