Minimate Pro Operator Manual
9
–29
a. Sensor Check:
Use the
Sensor Check
feature to determine if the attached sensors are connected,
oriented, and working properly, or to ensure that the sensors match the current settings
within the
View/Edit Setup File
menu if it has been set to
Before
or
After
an Event.
The results of the
Sensor Check
and the associated information for each sensor,
including
Serial Number
,
Calibration Date
, and
Calibration Location
, appear on the
Event Report
.
A
Sensor Check
may be setup to occur automatically, or by manual key-press:
(1)
Repeatedly press the
Right Arrow
key to set when a
Sensor Check
will occur,
After Event
or
Before Event
. You can also select
Disable
. If you select
Disable
, a
Sensor Check
can still be performed manually by pressing the
Sensor Check
key on the monitor. The manual
Sensor Check
is only allowed
when the unit is not in monitor mode.
If
Sensor Check
is set
Before Event
, the check is performed once and added to
all subsequent Event data. If the
Sensor Check
is set to
After Event
, then the
Sensor Check
is performed after each Event and added to each Events data.
(2)
Press
Enter
key to
Save
any changes, or the
Cancel
key to return to the
Preferences
menu, or
(3)
Press the
Sensor Check
key to perform a manual check while at the main menu.
NOTE:
The
Sensor Check
is not a calibration check and cannot replace the annual
calibration process.
Sensor Check
does not compare the measured result against an
external traceable reference sensor, nor does it test the entire electronic circuit that
is integral to the sensors response.
NOTE:
The sensors selected within the
Edit Current Setup
menu should match what is
attached to the unit.
Example:
If the
Pressure Sensor
Channel is
Off
and a
Linear Microphone
is
attached, or if it is
On
, and a
Linear Microphone
is not attached, the
Sensor
Check
will return the message, “
Sensor Detection Error – Attached Sensors do
not match setup
”.
When a
Sensor Check
error occurs, check the following for possible causes:
o
Was the sensor moved during the
Sensor Check
? This will result in a
failed reading.
o
Ensure that the orientation of the Geophone is correct, as per the given
application, such as a ceiling installation. Refer to the section in this manual,
Geophone Direct Hard Surface Mount Installation Instruction.