![Instantel Minimate Pro Series Operator'S Manual Download Page 41](http://html1.mh-extra.com/html/instantel/minimate-pro-series/minimate-pro-series_operators-manual_2069794041.webp)
Minimate Pro Operator Manual
7
–7
7.6.5. Using the Sound Level Microphone
1.
Sensor Check
is only available in Compliance Mode. Press
Sensor Check
on the
Minimate Pro
to ensure that the
sensor is properly attached. A successful
Sensor Check
will return results similar to this;
MicA
Pass
Freq= N/A
Amp=563 mv
NOTE:
The name of „MicA‟ is editable in the Blastware setup.
2.
Press
Enter
to continue.
3.
Press
Start Monitor
on the unit.
4.
If the unit triggers immediately you may need to increase the
Trigger Level
. Review the
peak of the previous event, and set the
Trigger Level
at least 3 dB above that value.
5.
Press
Start Monitor
.
6.
If the unit triggers immediately repeat the previous step otherwise clap your hands in
front of the microphone to ensure the unit will trigger.
7.
The system will now monitor and record data whenever the
Trigger Level
has been
exceeded.
7.6.6. View Compliance Events
1.
Once the monitor has recorded at least one event it will display the Lmax level of that event
on the display.
2.
To view other recorded data on the unit press
Down Arrow
from the main display and
highlight the desired event.
3.
Press
Right Arrow
three times to display the results of the
Pressure Channel
. The sound
data results that will be displayed are the Lmax, Lmin, L90, L10 and Leq values for the entire
recording time.
NOTE:
Refer to the Blastware Operator Manual for information about how to view, analyze
and/or export the recorded data for Compliance and Advanced mode software.
7.6.7. Technical Information for IEC 61672-1 Compliance
1.
The Sound Level Microphone is a Group Z, Class 1 instrument.
a.
Group Z refers to an instrument consisting of two separate components that must be
connected.
2.
Dependent on which Blastware software module that is used, Compliance or Advanced,
different data results may be achieved or settings applied, as shown below:
Specifications
Compliance Mode
Advanced Mode (Optional)
Weighting
A-Weight and C-Weight
Unfiltered
Frequency Range
Up to 20 kHz
Up to 20 kHz
Response
Fast 125 ms, Slow 1 s
N/A
Sensitivity
N/A
0.0038 V/Pa (Normal), 0.12 V/Pa
(Sensitive)
Resolution
0.05 dB (display limit 0.1dB)
0.03 Pa (Normal), 0.89 mPa
(Sensitive)