9
–14
Minimate Pro Operator Manual
(5)
Record Mode - Vibration Dose Value (VDV) [Optional]
The optional
Vibration Dose Value
(
VDV
)
Record Mode
is available on the unit with
the purchase of the
Instantel® Blastware Advanced Module
software. To order your
copy of the
Blastware Advanced Module
, please contact your
Instantel Dealer
Representative
. If you have a licensed copy of the
Blastware Advanced Module
,
follow the instructions in the
Blastware Operator Manual
for
Reloading the
Operating System
of the
Minimate Pro
, which will also load the
VDV
recording mode
into the unit if it is not already there.
The
VDV
monitoring option is used to measure and determine whole-body vibration
based on a
Frequency Weighting
file that is downloaded from
Blastware
. The
VDV
Record Mode
will not be allowed unless a
Frequency Weighting
file is downloaded to
the unit. Your geographic location and the type of vibration that you are monitoring will
determine which
Frequency Weighting
file that you choose to use during the monitoring
session – the frequency range of the data collected is from 1 to 80 Hz, therefore the
Instantel 1-80 Hz DIN Geophone
(Part No. 720A2003) is required for this type of
recording, as the sample rate is reduced to 512 s/s. Multiple
VDV
frequency weight files
may be downloaded to the monitor using the
Blastware Advanced Module
software.
Refer to the
Blastware Operator Manual
for more information on how to download
weight files to the unit.
While monitoring, the unit differentiates the velocity data into acceleration, applies the
frequency weighting and then calculates the
VDV
and
Root Mean Square
(
RMS
)
values. The unit displays the
VDV
and
RMS
values every ten seconds while recording,
and stores the final calculations as part of the file and Event Report.
There are two primary standards for determining and evaluating whole-body vibration, as
follows:
BS-6841
Measurement and evaluation of human exposure to whole-
body mechanical vibration and repeated shock.
ISO-2631
Evaluation of human exposure to whole body vibration.
These
Series IV
units are capable of continuously monitoring full waveform data for 24
hours; the daytime
VDV
monitoring is performed on the vertical channel (7:00 AM to
11:00 PM) while the nighttime
VDV
monitoring is performed on two horizontal channels
(11:00 PM to 7:00 AM) – the same or different Frequency Weight file can be use for
each of these monitoring periods.
NOTE:
The
VDV Record Mode
consumes more battery power than other record modes.
We recommend connecting the monitor to the supplied AC adapter or an external 12
VDC power supply.
(1)1.
To select the Vibration Dose Value Record Mode:
1.
Within the
Edit Current Setup
menu, press the
Up
or
Down
key to locate and highlight
the
Record Mode
menu option.
2.
Press the
Right Arrow
key to access the menu. Continue to press the
Right Arrow
key
until the
Record Mode
displays
Vibration Dose Value
. The following menu will be
displayed:
Summary of Contents for Minimate Pro Series
Page 1: ......
Page 12: ...2 6 Minimate Pro Operator Manual Notes...
Page 14: ...3 2 Minimate Pro Operator Manual Notes...
Page 16: ...4 2 Minimate Pro Operator Manual Notes...
Page 34: ...6 16 Minimate Pro Operator Manual...
Page 44: ...7 10 Minimate Pro Operator Manual Notes...