63
730
VIEW
9.10 Setting of D/A Output
The instrument can output voltage, current, active power, apparent power, reactive power, power factor, phase angle,
frequency, voltage peak, current peak and integral value with DC analog voltage of ±5VFS. The system allocates 12 item
channels for the output items, and each channel can select to output different measured values.
1. Press Utility key;
2. Select the "DA" function menu;
3. Confirm;
4. Select the fixed mode of DA output;
5. Confirm;
6. Exit and close the setting menu.
Follow the above steps to enter the DA output setting interface, and the instrument has configured two fixed modes for DA
output items. In addition, the user can also select custom output items.
The characters "DFIL-N", "DFLT-I" and "SEL" are displayed in the Display Frame C respectively. When "DFIL-N" and "DFLT-I"
are selected, it enters the data mode of fixed output; when "SEL" is selected, it enters the data mode of D/A custom
output.
D/A output mode
1)DFLT-N is the preset conventional measured value, and the output item of each channel is fixed as follows
NOTE : Among them, U4, I4 and P4 represent the measured values of the wiring group.
Channel
Item
CH1
U1
CH2
U2
CH3
U3
CH4
U4
Channel
Item
CH5
I1
CH6
I2
CH7
I3
CH8
I4
Channel
Item
CH9
P1
CH10
P2
CH11
P3
CH12
P4
9.10.1 Selection of fixed output mode
Operating steps