1. Press the Set/Reset key;
2. Select the integral D/A timer setting menu;
3. Confirm;
4. Set the hour (Display B);
5. Confirm;
6. Set the minute (Display C);
7. Confirm;
8. Set the second (Display D);
9. Confirm;
10. Exit and close the setting menu.
Follow the above steps 1~3 to enter the DA timer setting interface, and the setting cursor will be displayed in Display Bar
B. When setting with navigation keys, press "Enter" to confirm, switch the setting cursor to Display Bar C to set the minute,
confirm it again, and then switch the setting cursor to Display Bar D to set the second. After the hour, minute and second
are set, press "Enter" again. The instrument returns to the previous menu to display, and press "Hold/ESC" key to exit the
setting interface.
Setting range of integral D/A timer: 00:00:00~10000:00:00.
6.5 Integral Function Limitations
When the integral is in progress
The following settings cannot be made:
Measurement mode, wiring mode, measurement range, peak factor, synchronous source setting, input filter, data update
rate, number of harmonic measurement, operational formula of harmonic measurement distortion rate, PLL source setting,
integral mode, integral timer setting, integral start action and integral reset.
Functions cannot be executed:
Proportion function setting, average function setting, BIT operation, Math operation function, zero level compensation
function, configuration file saving and loading function.
When the integral is paused
The following settings cannot be made:
Measurement mode, wiring mode, measurement range, peak factor, synchronous source setting, input filter, data update
rate, number of harmonic measurement, operational formula of harmonic measurement distortion rate, PLL source setting,
integral mode, integral timer setting and integral stop action.
Functions cannot be executed:
Proportion function setting, average function setting, BIT operation, Math operation function, zero level compensation
function, configuration file saving and loading function.