46
VIEW 5 Pro
730
VIEW
8.1 Setting of GP-IB Bus
The remote control function of the instrument can be realized through GP-IB bus communication. When connecting to PC, it
is required to set the address of the instrument in GP-IB system, which is unique and can be set between 1~30.
Operating steps
1. Press Utility key;
2. Select the "coon" setting menu;
3. Confirm;
4. Select the GP-IB communication mode;
5. Confirm;
6. Set the address of GP-IB (1~30);
7. Confirm.
Follow the above steps 1~5 to enter the setting interface of GP-IB. Display Bars A, B and C fixedly display the characters
"UTIL", "CONN" and "GP-IB" respectively, Display Bar D is used to set the address bit, in which the address can be
switched with navigation (up and down) keys. After the setting is completed by pressing the "Enter" key, the display
returns to the main interface to display the measured value.
Select either GP-IB or RS-232.
⚠ ATTENTION
The instrument does not support setting the same GPIB address in GPIB cascade mode.