1. Press the Setup key;
2. Select the "avg" setting menu;
3. Confirm;
4. Select the average type;
5. Confirm;
6. Select the average type (exponential or linear);
7. Confirm;
8. Set the number of linear averages (8, 16, 32, 64) or the attenuation constant of exponential averages (8, 16, 32, 64);
9. Confirm;
10. Exit and close the setting menu.
It is operated according to the above steps to select and set the average type option "TYPE" in Display Bar C, and the
average type option "EP" or "LIN" is displayed in Display Bar C. It is operated in sequence according to the above two
paths to set the attenuation constant at exponential average and the average number at linear average respectively. The
attenuation constant and the average number are displayed in Display Bar D.
Description and limitations of average function
The averaging function is not available when the integral starts. Even if average is turned on, average will be automatically
turned off after the integral starts.
In harmonic measurement, only exponential average takes effect.
3.10 Scaling Function Setting
3.10.1 Turn scaling function on/off
Operating steps
1. Press the Setup key;
2. Select the "scale" menu;
3. Confirm;
4. Set the scaling function on/off;
5. Confirm;
6. Exit and close the setting menu.