18
VIEW 5 Pro
730
VIEW
Operating steps
Description of measurement mode
One of the following measurement modes can be selected for voltage and current measurement, and the default setting is
RMS.
When RMS mode is set, the true RMS of voltage and current are displayed, and the apparent power (S), reactive power (Q),
power factor (λ) and phase angle (Φ) are calculated and displayed in Urms*Irms.
When the Voltage Mean mode is set, the rectified average value of the voltage and current RMS is displayed, and the
apparent power (S), reactive power (Q), power factor (λ) and phase angle (Φ) are calculated and displayed in Umean*Irms.
When DC mode is set, DC voltage and current are displayed, and apparent power (S), reactive power (Q), power factor (λ) and
phase angle (Φ) are calculated and displayed in Udc*Idc.
1. Press the Setup key;
2. Select the "Mode" menu;
3. Confirm;
4. Select setting options;
5. Confirm;
6. Exit and close the setting menu.
The above operating steps are displayed. The measurement mode options are displayed in Column C. The options are
switched through the navigation (up and down) key. When the measurement mode options to be set appear, the "Enter"
key is clicked to complete the setting. The indicator light of the corresponding setting item on the panel is on. The
instrument returns to the superior menu for display. Pressing the "Hold/ESC" key can exit the setting interface.
3.1 Measurement Mode Setting
Indicator light
Voltage
Current
RMS
True RMS
True RMS
VOLTAGE MEAN
Rectified average calibrated to RMS
Rectified average calibrated to RMS
DC
Simple average
Simple average
Chapter III Setting of Measurement Conditions