12
VIEW 5 Pro
730
VIEW
2.3 Turn the Power On / Off
Operation when turning on the power
●
As shown in the schematic diagram of the front panel in Section 1.1, the front power switch key and the rocker switch
on the back of the instrument together form a power double switch design.
●
When the rocker switch on the back is not closed, and the indicator light of the power key is not on, the instrument is
not charged.
●
After the rocker switch on the back is closed, the instrument is powered on, the indicator light of the power key is red,
and the instrument is in standby state. At this time of pressing the power switch key, the instrument starts, the indicator
light of the power key is green, and the instrument enters the running state.
●
When the instrument is running, the power switch key is pressed for up to 2 seconds, the instrument stops running and
enters the standby state, the digital tube is turned off, and the indicator light of the power key turns red at the same time.
Precautions when making accurate measurements
●
The power switch is turned on to enable the instrument preheat for more than 30 minutes.
●
Please set it to zero after preheating.
Operation when turning off power
●
Please turn off the power switch first and then the rocker switch when the instrument needs to be turned off after use.
When the instrument is completely uncharged, it is regarded as normal shutdown.
●
If the instrument is abnormally powered off and shut down, the instrument starts automatically when it is powered on
again. Please try to use the normal shutdown process to close the equipment. Please wait for more than 2S to power on
and start the instrument again in case of abnormal shutdown.
2.4 Measurement Circuit Connecting Direct Input
The measurement cable from the measurement circuit is connected directly to the voltage or current input terminal.
●
Connect voltage input terminal
For a safety banana jack with a terminal of 4mm, the safety connector (the wire is not exposed) is inserted into the voltage
input terminal.
●
Connect current input terminal
It is not allowed to touch the current sensor input interface when introducing the voltage of the measurement circuit
into the current input terminal. Because these ports are electrically connected internally, they are very dangerous.
⚠ CAUTION
●
Please make sure that there is no foreign matter on the contact surface between the current input terminal and the
crimp terminal when connecting.
●
It is necessary to regularly check whether the knob of the current input terminal is loose, and whether there is foreign
matter on the contact surface between the current input terminal and the crimp terminal.
Voltage input end
Current input end