23
730
VIEW
1. Press the Setup key;
2. Select the "f.filt" setting menu;
3. Confirm;
4. Set the frequency filter switch;
5. Confirm;
6. Exit and close the setting menu.
It is operated according to the above steps to enter the switch setting interface of the frequency filter. The switch
options of the frequency filter are located in Display Bar C. The switch option "OFF" or "ON" can be switched through
the navigation (up and down) key, and pressing the "Enter" key can complete the setting. The instrument returns to the
superior menu to display, and pressing the "Hold/ESC" key can exit the setting interface.
3.8 Line Filter Setting
The line filter is inserted in the voltage and current measurement circuit, thus it will directly affect the measurement of voltage, current
and power. When the line filter is turned on, the measured value does not contain high frequency components, thus the noise from the
frequency converter or distorted waveform can be removed to measure the voltage, current and power.
Operating steps
1. Press the Setup key;
2. Select the "l.filt" setting menu;
3. Confirm;
4. Setting the line filter switch;
5. Confirm;
6. Exit and close the setting menu.
Description
ON is set to turn on the line filter. When it is turned on, the default cutoff frequency is 500Hz, and the "Line" indicator light
on the panel is on.
Setting OFF indicates that the line filter is disabled, and the indicator light is off.
Description
ON is set to turn on the frequency filter. When it is turned on, the default cutoff frequency is 500Hz, and the "Freq"
indicator light on the panel is on.
Setting OFF indicates that the frequency filter is disabled, and the indicator light is off.