14 |
37720-03F
dissolved and incorrectly deployed the esophageal retention
flange inside the puncture tract rather than inside the lumen of
the esophagus, considerable resistance to device rotation will
be felt.
A second method for confirming esophageal retention flange
deployment is to utilize direct flexible endoscopy confirmation
of deployment.
If the assessment methods described above fail to provide
satisfactory verification of correct esophageal retention flange
deployment, remove the device, dilate and remeasure the
puncture tract, and repeat the device insertion and confirmation
process.
Once deployment of the esophageal retention flange has been
confirmed, detach the neck strap from the tracheal retention
flange (diagram 1b) at the area of reduced neck strap width,
where it meets the tracheal flange, by cutting carefully with
scissors (diagram 6).
Note: if you are uncertain that the esophageal retention flange
has properly unfolded, and the device is in place inside the
esophageal lumen, after using the steps described above, do
not detach the neck strap from the device. Securely tape the
neck strap to the neck and observe the patient for a few days for
retention ability.
Removal of the Device
The Indwelling Occluder is not a permanent device and requires
replacement periodically. Many clinicians recommend routine
follow-up evaluations not longer than 6 months. InHealth
Technologies makes no warranty, either expressed or implied, as
to the lifetime of the Indwelling TEP Occluder. The device lifetime
may vary with individual use and biological conditions.
Removal of the Indwelling Occluder should only be done by
grasping the outer tracheal flange of the device securely with
a locking hemostat. Pull gently and firmly until the device is fully
withdrawn.
Insert a Dilator-Sizer of the appropriate diameter, and tape it in
position for five minutes prior to inserting a new Indwelling Occluder.
Summary of Contents for Blom-Singer Indwelling TEP Occluder
Page 1: ...37720 03F R3 Blom Singer Indwelling TEP Occluder MEDICAL PROFESSIONAL Instructions For Use...
Page 4: ...4 37720 03F...
Page 16: ...16 37720 03F...
Page 28: ...28 37720 03F...
Page 41: ...ObturatorTEPPermanentBlom Singer 37720 03F 41...
Page 53: ...Blom Singer TrajniTEPOkluder 37720 03F 53...
Page 65: ...Blom Singer TrajniTEPOkluder 37720 03F 65...
Page 77: ...NotranjeZapiraloBlom Singer TEP 37720 03F 77...
Page 78: ...78 37720 03F BLOM SINGER Blom Singer 1 2 Blom Singer InHealth Blom Singer Blom Singer...
Page 80: ...80 37720 03F Blom Singer InHealth Technologies Blom Singer Vaseline...
Page 81: ...O Blom Singer 37720 03F 81 Blom Singer 24...
Page 82: ...82 37720 03F 6 InHealth Technologies...
Page 83: ...O Blom Singer 37720 03F 83 Blom Singer Blom Singer Blom Singer Blom Singer 1 1 1b 1...
Page 86: ...86 37720 03F Blom Singer a 4a 4b 4c 4d 4e...
Page 87: ...O Blom Singer 37720 03F 87 1 3 2 3 5 Vaseline 4 1...
Page 88: ...88 37720 03F 5 1c...
Page 89: ...O Blom Singer 37720 03F 89 Eric D Blom 3 1b 6...
Page 90: ...90 37720 03F 6 InHealth Technologies...
Page 92: ...92 37720 03F...
Page 93: ...37720 03F 93 DIAGRAMS Diyagramlar Diagrame Dijagrami Diagrami 2 2a 1 a c b...
Page 94: ...94 37720 03F 4a 3 4b 4c 4e 4d...
Page 95: ...37720 03F 95 5 6...
Page 98: ...98 37720 03F...
Page 99: ...37720 03F 99...