7 | Warning and error messages
INFICON
68 / 90
HLD6000-Operating-instructions-kinb43en1-05-(2107)
No.
Message
Possible error sources
Troubleshooting
W543
Flow in the sniffer line
is too low
• Filter in the sniffer tip is clogged
• Sniffer tip is clogged or
defective
• Internal filter of sniffer line
blocked (only PLUS sniffer line)
• Change the filters, see also
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
W544
Valve does not toggle Internal defect of the sniffer line.
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
W545
Flow at measuring
line too low
• The filter in the sniffer tip is
clogged.
• Sniffer tip is clogged or
defective.
• Internal filter of sniffer line
blocked (only PLUS sniffer line)
• Replace the filters
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
W546
Leak at measuring
line
• The filter in the sniffer tip is
clogged.
• Leak or defect in the sniffer tip
or the sniffer line.
• Replace the filters.
• Check the plug and screw
connections.
• Check the connection of the
sniffer line with the basic unit
(disconnect and reconnect; if
possible try a different sniffer
line).
• If the problem persists, contact
the INFICON customer service.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......