5 Installation
5.1 Setup
DANGER
Danger to wearers of pacemakers or implants
The device contains magnets. Pacemakers and other implants can be influenced in
their function.
► Keep a distance of at least 10 cm between the device and the implant.
WARNING
Danger from moisture and electricity
Moisture entering the device can lead to personal injury due to electric shocks as well
as damage to property due to short circuiting.
► Only operate the device in a dry environment.
► Operate the device away from sources of liquid and moisture.
NOTICE
Property damage due to overheated device
The device heats up during operation and can overheat without sufficient ventilation.
► Please note the technical data, see “Technical specifications [
21]“.
► Ensure sufficient ventilation, especially on the ventilation slots on the left and right
of the device: At least 20 cm of free space on the sides, at least 10 cm in the front
and rear.
► Keep heat sources away from the device.
► Do not expose the device to direct sunlight.
5.2 Sniffer line
5.2.1 Connecting the sniffer line
Connect the sniffer line before you start up the device.
If the sniffer line is not connected then the device reports an error.
If you remove the sniffer line during operation, the device also reports an error.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......