5.3.2 Using a water conservation sniffer tip
If there is a risk of sucking in liquids then a water conservation tip should be used, see
“Accessories and spare parts [
80]“.
You can use the water conservation sniffer tip to check test objects with low surface
moisture, e.g. condensation moisture, for leaks.
► Screw the water conservation sniffer tip to the end of the filter holder.
Fig. 8: Water protection sniffer tip installed
NOTICE
Risk of short circuit
Sucked in liquid can destroy the device.
► Do not suck up liquids with the instrument.
► Do not hold the sniffer line with the sensor pointing up since moisture will flow into
the sensor as a result. If liquid has been spilled in the direction of the sensor, hold
the sniffer line with the tip pointing down and run the device for about 10 minutes.
With the water conservation sniffer tip installed, you cannot calibrate with the internal
COOL-Check.
5.3.3 Using flexible sniffer tip
If you are examining test objects which have already been packed or test objects with
areas that are difficult to access then an extended sniffer tip can be used, see
"Accessories and spare parts [
In addition to the rigid sniffer tip what comes delivered with the HLD6000 you can also
use a 400 mm long flexible tip. By bending the flexible tip accordingly, hard-to-reach
areas can also be accessed.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......