INFICON
Operation | 6
HLD6000-Operating-instructions-kinb43en1-05-(2107)
51 / 90
3
Switch to measurement screen
.
4
For external calibration, select
.
Fig. 13: Display for external calibration
5
If you wish to change the displayed leakage rate (in this example: 12.3 g/a) then
this can be done after pressing a finger on this value. For more information see
"Assembly of the touchscreen [
6
Hold the sniffer tip to the opening of the external test leak and start the
calibration by pressing the green button on the touch screen or alternatively by
pressing the button on the sniffer probe.
7
Hold the sniffer line still and straight as long as the device is calibrating.
ð
The display shows individual calibration phases and informs you whether or not
calibration was successful.
6.3.6.4 Checking the calibration with an internal COOL-Check
The calibration can also be checked without changing calibration values. You can find
out whether calibration is necessary.
1
Switch to measurement display.
2
Lead the sniffer tip into the calibration opening on the front of the basic unit
without pressing the button on the sniffer probe.
3
Hold the sniffer line still and straight as long as the device is checking the
calibration.
ð
A message appears stating whether the calibration is OK or whether the
device requires recalibration.
ð
When the corresponding message appears on the measurement screen then
press the button on the sniffer probe to calibrate.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......