INFICON
Operation | 6
HLD6000-Operating-instructions-kinb43en1-05-(2107)
47 / 90
Preset gases
Select preset gas
1
> Gas
2
Set the desired gas.
ð
If you have internally calibrated a preset gas and are switching over to
another preset gas then recalibration is not required.
ð
If you have externally calibrated a preset gas and are switching over to
another preset gas then the device prompts you to calibrate.
User-defined gases
Enter name for user defined gas
You can enter a maximum of 3 additional gases of your choice, provided that these
can be verified by the device. You can obtain more information about this from
INFICON by request.
1
> Setup > Advanced setup > SMART user gases
2
Enter the desired name under <Name user gas 1> or <Name user gas 2> or
<Name User gas 3>.
ð
You can set the values for the user defined gas.
Set values for the user defined gas
Option 1: Calibrate the device with the desired gas using an external calibration leak.
1
> Gas
2
Set the desired user-defined gas.
ð
The calibration factor should be set to "0" in the settings window under
"Factor user gas".
3
Then calibration with the external calibration leak can be carried out, see
“Calibrate with external calibration leak [
50]“.
Option 2: Calibrate the device with the desired gas without an external calibration leak.
ü
You have an internal COOL-Check as the basis for calibration with the refrigerant
R134a.
ü
You know the calibration factor for automatic deviation correction. For most gases
the required calibration factors from INFICON can be used.
1
> Gas
2
Set the desired user-defined gas.
3
In the setting window under "Factor user gas" enter the calibration factor which
you received from INFICON.
ð
When carrying out the unit calculation of user gases in mbar l/s and Pa m
3
/s
an average molecular weight of 96 is assumed. If this is not accurate enough
then calibrate externally using a test leakage value in the corresponding unit.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......