6 Operation
6.1 Switch ON
Connect a sniffer line and switch on the device.
The device will start up and, after a short time, display "Reading data".
Fig. 11: The device starts up
► In order to switch to English from another language, press "EN" on the depicted
window.
After the run-up, the device will measure the leakage rate on the sniffer line. There is
no separate Start function.
HLD6000 with one PLUS sniffer line
The device must typically be in operation for at least 30 minutes to achieve all
measuring and filtering characteristics. The measurement sensitivity may change
within the first 30 minutes.
► Do not calibrate or measure until 30 minutes after switching on the device.
1
Make basic settings, see "Basic settings [
2
Make adjustments to settings for the measuring process and calibrate the
device , see “Settings for the measurements [
45]“.
If the COOL-Check calibration leak is not located in the device, an acoustic warning
signal sounds and the 163 warning message is issued during the initial start-up.
If calibration is required, the warning 630 "Calibration request" is issued, see
"Measurement display elements [
18]".
► To close this warning message, press the
button.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......