6.10 Resetting to factory settings
You can reset the device to the factory settings, separately for the settings of the
operating unit, the basic unit or the parameter access level.
Loss of settings and access rights
After resetting to factory settings only the manufacturer factory settings are in the
memory of the device. The software of the device is not reset when restoring the
factory settings.
1
> Parameter > Reset
2
Make your choice:
ð
If you want to reset operating unit settings such as display settings, press the
"Reset" button next to the "Operating unit settings". See also “Factory settings
ð
For example, if you want to reset measurement settings such as setpoints,
press the "Reset" button next to the "Basic unit settings". See also “Factory
ð
If you want to restore the factory settings for access, press the "Reset" button
next to "Parameter access level". See also “Factory settings [
23]“, Table 2.
ð
To reset the device to the factory settings, press all "Reset" buttons one after
the other.
See also
2
6.11 Updating the software
Software updates from INFICON are installed with the aid of a USB stick. To the
update function of the device:
►
> Update
An update is possible,
• if one or several updates are available on the USB stick, but only one update per
type at most (basic unit, operating unit, sniffer line, I/O module),
• if in the case of the "sniffer line" or the "I/O module" these parts are also
connected free of disturbances and have an update function.
The corresponding buttons in the update menu such as "Basic Unit", "Operating Unit",
"Sniffer Line" and "I/O Module" are active and can be activated individually.
Summary of Contents for HLD6000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 79: ...INFICON Decommissioning 9 HLD6000 Operating instructions kinb43en1 05 2107 79 90 ...
Page 85: ...INFICON Appendix 10 HLD6000 Operating instructions kinb43en1 05 2107 85 90 10 4 RoHS ...
Page 88: ...Index INFICON 88 90 HLD6000 Operating instructions kinb43en1 05 2107 ...
Page 89: ......
Page 90: ......