5.3 Sniffer tip
5.3.1 Replacing the sniffer tip
NOTICE
Material damage from pollution
Particles in the air intake can destroy the sniffer line.
► Always switch off the device before every installation!
► Make sure that no particles enter into the air intake when replacing sniffer tip parts.
Fig. 7: Sniffer tip
1 Filter block
2 Guide pin
3 Cap nut
4 Sniffer tip
5 Filter holder
1
Switch the instrument off.
2
Loosen the cap nut until the sniffer tip can be pulled out.
3
Insert the new sniffer tip into the sniffer probe in such a way that the guide pin
runs into the groove of the sniffer probe opening.
4
Tighten the cap nut.
5
Calibrate the device see “Time and type of calibration [
48]“.
ð
Further sniffer tips see “Accessories and spare parts [
80]".