NanoPhotometer
®
N120/
NP80/N60/N50/C40
User Manual Version 4.3.1
145
8.
M
AINTENANCE
M
AINTENANCE
F
REE
T
ECHNOLOGY
The NanoPhotometer
®
technology is maintenance free. Regular maintenance and calibration
is not necessary.
For facilities that are working according to national as well as international guidelines and
standards including: Good Laboratory Practice (GLP), Good Manufacturing Practice (GMP),
or ISO9000-9004; the proper performance of the spectrophotometer has to be tested and
proven on a regular basis with individually set intervals. Implen provides certified
NanoPhotometer
®
secondary standards as an optional accessory. These NanoPhotometer
®
Didymium glass filters and standard solution sets are suitable for the control and
documentation of the wavelength accuracy and the photometric accuracy of your system.
IQ/OQ documentation is also available. Please contact your local Implen office or an
authorized Implen partner for further information.
Note
: Further information on the secondary standards can be downloaded from the Implen
website in the download area - quality control (
Support agreements that help to fulfill the demands of regulatory guidelines concerning
GLP/GMP include: calibration certification using filters traceable to international standards
(during production and quality control), certified engineers and calibrated test equipment,
approved to ISO 9001 standard, automatic self-diagnostic calibration test during start of the
NanoPhotometer
®
, result is documented in each data output file, and possibility to save a
Implen Document Source (IDS) file (no data manipulation possible).
R
EPLACEMENT
P
ARTS
Lamp Replacement
The NanoPhotometer
®
is equipped with a xenon flash lamp with a lifetime of 10
9
flashes (up
to 10 years). This lamp should not need replacement for several years. In the unlikely event
the lamp does need to be replaced, this should be done by the manufacturer or a certified
service engineer from your supplier.
Battery Replacement
The optional battery pack can only be assembled or replaced by the manufacturer or a
certified service engineer from your supplier.
Note
: Danger of explosion if battery is incorrectly replaced. Replacement only with the same
or equivalent type recommended by the manufacturer: Replacement needs to be done by a
certified service engineer
Note:
Battery operation time is depending on display usage. Activating the screen saver
video may reduce the battery operation time.
Note
: Recharging of deep discharged batteries is not possible they need to be replaced by
new ones.