
NanoPhotometer
®
N120/
NP80/N60/N50/C40
User Manual Version 4.3.1
102
If it is desired to heat the sample to 37°C use the toggle switch to turn on cell holder heater.
When the cuvette holder has reached 37°C the toggle color changes to green.
Note
: Only available for cuvette applications (NP80 and C40).
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different
lids see page 52 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is shown.
Note:
A new Blank is recommended when changing the dilution lid.
4. Enter desired wavelength (
) to be measured. It is possible to measure up to 20
wavelengths simultaneously. More wavelength (
) options can be added by selecting the
Add Wavelength button. Added wavelength can be deleted with
5. Baseline correction is set off by default. Enabling the baseline corrections shows a list with
different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm).