NanoPhotometer
®
N120/
NP80/N60/N50/C40
User Manual Version 4.3.1
141
Update procedure:
1. Unzip the installation file to a USB flash drive into the root folder
2. Insert in the USB flash drive to the USB port of the NanoPhotometer
®
3. Select Assistance/Software Maintenance
4. Push
on “Update” and wait until the NanoPhotometer
®
reboots
Note
: Once you have updated the firmware of the NanoPhotometer
®
, please update the
client software (PC/Mac, tablet, smartphone).
C
REATE
L
OG
F
ILE
To create a log file insert a USB flash drive to the USB port of the NanoPhotometer
®
. Push
on “Create Log File”. The log (NPOS.log) file will be saved on the root folder of the USB
flash drive.
F
ACTORY
R
ESET
There is an option to reset the instrument to factory settings. By selecting the reset button a
window will open that says “Reset the NanoPhotometer
®
?” Selecting the cancel button will
close the window without changing the settings and selecting reset will open a window that
will ask again ”Reset the NanoPhotometer
®
to factory settings? All data, stored methods and
settings will be lost.” If it is confirmed the factory settings will be restored.
Note:
All stored methods, settings and data on the NanoPhotometer
®
will be deleted if the
reset option is executed.
C
REATE
B
ACKUP
Option to create backup file to recover data and settings of the NanoPhotometer
®
at a later
date. To create a backup file insert a USB flash drive to the USB port of the
NanoPhotometer
®
. Push
on “Create Backup”. The backup (NPOS.bak) file will be saved into
the root folder of the USB flash drive.
Recovery of the NanoPhotometer
®
is only possible via the Implen support (remote access).
In the case a NanoPhotometer
®
recovery is necessary please get in touch with the Implen
support team (