NanoPhotometer
®
N120/
NP80/N60/N50/C40
User Manual Version 4.3.1
138
port.
initialization test. Restart the NanoPhotometer
®
with the cuvette port empty. If the message still
appears please create a log file and send it to
Implen Customer Support.
NanoVolume head not clean! Please clean
sample windows thoroughly.
Clean the sample window on pedestal and the
mirror in the lid arm with 70% EtOH and distilled
water and then reboot the NanoPhotometer
®
. If
the message still appears please create a log file
and send it to Implen Customer Support.
No connection possible
Connection is lost between control devices and
NanoPhotometer
®
. Check the connection. If the
message still appears please create a log file
and send it to Implen Customer Support.
Server Not Available
Internal software communication problem during
an open measurement session. Reboot the
NanoPhotometer
®
. If the message still appears
please create a log file and send it to Implen
Customer Support.
Server Side Failure
Internal software communication problem during
an open measurement session. Reboot the
NanoPhotometer
®
. If the message still appears
please create a log file and send it to Implen
Customer Support.
Software update failed. Please contact Implen
customer support.
Make sure that the sample name is NPOS.bin
without any additions or extensions like from
multiple downloads, otherwise rename file to be
NPOS.bin exactly. NPOS.bin file must be in the
root folder of the USB flash drive. Try the update
a second time, if it still fails please create a log
file and send it to Implen Customer Support.
Stored method/result was created for unavailable
mode NanoVolume/Cuvette/Submicroliter Cell!
Message appears if stored results or a stored
method made on another NanoPhotometer
®
type
is opened with an unavailable mode like Cuvette
or NanoVolume e.g. data or methods made in
cuvette mode and opened on a N60
(NanoVolume only).
S
YSTEM
F
REEZE
In case of harsh ESD impulses it could happen that the device does freeze or switch-off.
Please wait first about 30 sec to see if the system recovers automatically itself. If there is no
reaction please press the power key at the instrument rear 4 sec to switch-off and then again
shortly to restart the system.