C O M M A N D R E F E R E N C E
Command Reference
74
LDC-3700C Series
C H A P T E R
4
The LASer:CALMD? query returns the value of the laser's photodiode feedback
responsivity (CAL PD parameter) setting.
Notes
If this value is 0, the LDC-3700C Series Laser Diode Controller will be set to operate in
constant I
PD
mode, and the I
PD
set point value will be in effect. If this value is non-zero, the
LDC-3700C Series Laser Diode Controller will be set to operate in constant P
PD
mode,
and the P
PD
set point value will be in effect.
If this value is 0, the front panel P
PD
will display "-.-", and no P
PD
value can be calculated.
Examples
"LASER:CALMD?" -response: 0, means the LDC-3700C Series Laser Diode Controller
is set for operation in constant I
PD
mode (if P mode is also selected).
"Las:Calmd?" -response: 1.1, means the LDC-3700C Series Laser Diode Controller is
set for operation in constant P
PD
mode (if P mode is also selected) and the responsivity
is set to 1.1
A/mW. 1.1
A of photodiode feedback current represents 1 mW of optical
power.
The LAS:CAL:CONST? query returns a stored calibration value. It is intended for factory
use only.
The LASer:CAL:LDI command is used to enter the LASER current set point,
measurement, and limit (in low bandwidth mode) calibration mode.
Parameters
None.
Notes
Since the limit circuit is the same for both high and low bandwidth modes, it is only
calibrated when low bandwidth mode is selected.
After this command is issued, the LDC-3700C Series Laser Diode Controller will allow
calibration of the current set point, measurement, and limit (if low bandwidth mode is
selected). This procedure is outlined in Chapter 6.
The calibration defaults to the selected bandwidth setting. Therefore, it is necessary to
select the desired bandwidth (I or I
HBW
), and turn the LASER output on before performing
the calibration for that bandwidth.
Calibration is performed at the current set point, wherever it is set. If the LASER output is
not on, E 517 is displayed.
In remote operation, the LASer:CAL:LDI? query may be used to determine if the LDC-
3700C Series Laser Diode Controller is ready for the user to enter a value.
Examples
"Las:CAL:LdI" -action: the LDC-3700C enters calibration mode for LASER current.
"Las:Cal:LDI" -action: the LDC-3700C enters calibration mode for L
LASer:CAL
MD?
PD?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
LAS:CAL:CONST?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
LASer:CAL:
LDI
I
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Summary of Contents for LDC-3700C Series
Page 2: ......
Page 8: ...L I S T O F FIG U R E S vi LDC 3700C Series...
Page 10: ...L I S T O F TA B L E S viii LDC 3700C Series...
Page 78: ...RE M O T E O PE R A T I ON Command Syntax 62 LDC 3700C Series C H A P T E R 3...
Page 150: ...C O M M A N D RE F E R E N C E Command Reference 134 LDC 3700C Series C H A P T E R 4...